Leica CM3050S – Cryostat

5.5.2 Freezing specimens onto specimen

discs with 90 ° prism (Instruments w/

specimen cooling only)

In instruments with specimen cooling, the 90°

prism, which is inserted into the specimen head,

is an additional option to freeze specimens at

maximum temperature .

Installing the 90° prism
• Loosen screw (3) at specimen head.
• Insert pin (1) of prism completely into recep-

tacle (2) in the specimen head.

Important: the entire rear surface of the

prism must fit tightly against the specimen

head surface!

• Retighten screw (3) at specimen head.
The prism now forms a horizontal surface for

quick specimen freezing.
Freezing specimens with the 90° prism
• Apply sufficient amount of embedding me-

dium onto ambient temperature specimen

disc, add specimen and orient.

• Insert specimen disc with specimen into the

receptacle (4) of the prism.

• Press

to activate menu item ‘Fast Object

Freezing’ and activate function by pressig

(see chapter 5.3.4 ‘Menu functions’ item 4).

5. Operation

The specimen head is now automatically

cooled down to the lowest temperature (-50 °C).
• Once the specimen is frozen, remove the

disc from the prism and store it in the quick-

freeze shelf until inserting it into the speci-

men head.

• Press

again to activate menu item ‘Fast

Object Freezing’ and deactivate the func-

tion by pressing

once again (see chap-

ter 5.3.4 ‘Menu functions’ item 4).

The specimen head temperature now automati-

cally returns to the previously selected set val-

ue.
• Loosen screw (3) at specimen head and re-

move prism (1).

• Store prism inside the cryochamber.

5.5.3 Optional accessories for freezing

specimens onto specimen discs and

for removing frozen specimens

Mobile heat extractor
In addition to the stationary heat extractor (part

of standard delivery), a mobile heat extractor

is available as an optional accessory - see Ap-

pendix ‘Ordering information’ - accessories ):

• Store heat extractor in chamber to precool.

• Place onto specimen in quick-freeze shelf.

• Remove as soon as specimen is frozen.