3.2 Unpacking and installation

Unpacking instructions are always located in a

transparent protective envelope on the outside

of the instrument shipping crate.

Make sure to read and follow all safe-

ty instructions provided in chapter

2.3.1 ‘Transport’ and on the unpacking

instructions!

3.2.1 Repacking

We recommend to keep the original shipping

crate and the unpacking instructions for the

Leica CM3050 S.
For repacking, proceed as per unpacking in-

structions, in reverse order.

Available models
• Basic instrument

- with sectioning motor

- without specimen cooling

• Basic instrument

- without sectioning motor

- with specimen cooling

• Basic instrument

- with sectioning motor

- with specimen cooling

3.1 Site requirements

Make sure to read and follow all safe-

ty instructions in chapter 2.3.2 ‘Site re-

quirements’!

3.1.1 General site requirements

• No direct sunlight.
• Electrical power supply within distance

(length of power cord = approx. 4 meters -

do not use extension cords!).

• No draft (caused by air conditioning etc.).
• Even floor surface.
• Practically vibration-free floor.
• Handwheel easily accessible.
• Room temperature constantly below +22 °C.
• Relative humidity of air maximum 60 %.

High ambient temperature and/or high

air humidity negatively affect instru-

ment cooling performance!

3.1.2 Electrical connections

Make sure to read and follow all safe-

ty instructions in chapter 2.3.3 ‘Electri-

cal connections’!