5. Operation

5.7.2 Manual chamber defrost cycle

In addition to the automatic defrost cycle, a

manual chamber defrost cycle can be activat-

ed when necessary.
• Press

until menu item ‘Set chamber

temperature’ is displayed:

• Press

. A brief sound signals that the

manual chamber defrost cycle has started:

The manual defrost cycle ends either when the

selected defrost cycle duration is over or when

a chamber temperature of -5 °C is reached.

To abort the manual defrost cycle:
• Go back to menu item ‘Set chamber temper-

ature’ and press

again.

Instruments with specimen cooling:

When defrosting starts, to avoid icing the speci-

men head is automatically adjusted to a tem-

perature between -10 °C and -5 °C, i.e. the set

temperature value is deactivated temporar-

ily. (same as with automatic defrost cycle - see

chapter 5.7.1 ‘Automatic chamber defrost cycle’)
10 seconds after the manual chamber de-

frost cycle has been completed, the specimen

head refrigeration is switched back on, cooling

down to the set value.

5.7.3 Defrosting the specimen head

(instruments with specimen cooling

only)

To activate specimen head defrosting:
• Press

until menu item ‘Set specimen

temperature’ is displayed:

• Press

. A brief sound signals that the

manual chamber defrost cycle has started:

Specimen head temperature increases to be-

tween +20 °C and +30 °C for a duration of

10 minutes, after which it returns to the set

value.
To abort specimen head defrosting:
• Press

until menu item ‘Set specimen

temperature’ is displayed and press

once more.

S E T T E M P C T - 3 0 ° C

-

S E T T E M P O T - 3 5 ° C

 

S E T T E M P C T - 3 0 ° C

-

S E T T E M P O T - 3 5 ° C