4 installation, Appendix – Leica Biosystems CM3050 S User Manual
Page 60
60
Instructions for Use V 1.7 RevC - 08/2014
9. Appendix
9.1.4 Installation
• Open the shipping box and remove all packing material including the Sty-
rofoam pieces.
• Take the height adjustment device out of the box.
• At the place of installation, place the U-frame on the floor with the open
side facing yourself. Adjust the U-frame parallel to the wall, maintaining a
distance of approx. 10 cm between wall and U-frame.
• Push the (up) button to lower the height adjustment device to the lowest
possible position.
• Plug the height adjustment device into the
mains socket. The unit is now ready for use.
• In order to be able to move the instrument,
use a fork wrench and unscrew the roll-
ers on which the Leica CM3050 S moves by
about 2 cm.
• Roll the cryostat onto the U-frame as far as it
will go. The rear of the cryostat housing has
to fit closely to the rear bar of the U-frame.
• Roll the cryostat onto the U-frame as far as it
will go. The rear of the cryostat housing has
to fit closely to the rear bar of the U-frame.
When placing the instrument onto the
U-frame, make sure the power cord
does not get caught and thus crushed
between U-frame and cryostat housing.