2.3.9 Infectious/radioactive material

• Danger!

Use caution when working with potentially

infectious specimens:

-->Risk of infection!
• When working with potentially infectious /

radioactive specimens:
• Wear protective clothes (gloves, protec-

tive boots, mask, lab coat), in compliance

with radiation safety regulations and/or

in-house regulations on handling infec-

tious / radioactive material.

• When working with radioactive specimens:

• Comply with applicable radiation safety

regulations!

• Dispose of radioactive specimen waste

according to applicable regulations.

2.3.10 Disinfection and cleaning

• Prior to disinfection, switch the instrument

off and unplug it from mains.

• For removal of the microtome from the cryo-

chamber see chapter 2.3.11 ‘Removing the

microtome’.

• For disinfection, wear protective gear:

(gloves, mask, lab coat etc.)!

• For disinfection, only use alcohol-based dis-

infectants!

• Do not use solvents (xylene, acetone etc.)

for cleaning or disinfection!

• Do not spray disinfectants into the evapora-

tor!

--> Risk of icing!

• Explosion hazard when working with alco-

hol: Make sure the premises are appropri-

ately ventilated!

• When using disinfectants and detergents,

comply with all safety instructions supplied

by the manufacturer of the product!

• Dispose of waste liquids from disinfection/

cleaning as well as of sectioning waste ac-

cording to applicable regulations on dispos-

al of special category waste!

• Disinfected accessories must be thoroughly

dry when reinserting them into the chamber!

--> Risk of icing!
• Make sure the chamber is completely dry

before switching the instrument back on:

--> Explosion hazard through alcohol vapors!

--> Risk of icing!