3 the footswitch, 5 prior to switching on the instrument, Prior to switching on the instrument – Leica Biosystems CM3050 S User Manual

Page 20: Installation

Advertising
background image

20

Instructions for Use V 1.7 RevC - 08/2014

3.4.3 The footswitch

Function
The footswitch performs the same task as the

RUN/STOP and RUN/ENABLE keys (activating/

deactivating motorized sectioning / trimming).

In addition, the footswitch can be used to acti-

vate the emergency stop function.

Models with footswitch:
All instruments with sectioning motor.

Important note:

In all instrument models that are deliv-

ered with footswitch, the footswitch

must be installed! - Otherwise the in-

struments are not functional.

After transport, observe a waiting pe-

riod of at least 4 hours before turning

the instrument on! - See also safety in-

structions in chapter 2.3.1 ‘Transport’.

Have you observed all safety instruc-

tions in chapters 2.3.2 ‘Site require-

ments’ and 2.3.3 ‘Electrical connec-

tions’?

If not:

--> Please read chapters 2.3.2 and

2.3.3!

• Insert mains plug into wall outlet.
• Continue with chapter 5.1 ‘Setting up the in-

strument’.

3. Installation

Connecting the footswitch
• Insert footswitch into port (1) and secure.

1

3.5 Prior to switching on the instrument