3 display indications when switching on, Operation – Leica Biosystems CM3050 S User Manual

Page 28

Advertising
background image

28

Instructions for Use V 1.7 RevC - 08/2014

5. Operation

Meaning of indication:

Operator activity:

yes

no

no

Acknowledge indication by
pressing

once briefly.

Initialization phase 1

Initialization phase 2

Software version

Selected display language

Sectioning motor speed

level ‘slow’ / (fast)

---

---

---

Select language via

keys

Select speed level via

keys

5.3.3 Display indications when switching on

Standard indication ‘locked’ is displayed:
(Display background dark)
To unlock: press

until the display back-

ground lights up:

yes

Temperature prior to

switching off < 0 °C

If this indication is displayed: Do not switch

on the instrument!

Prior to switching on, both microtome and ac-

cessories must be completely dry!

Remove microtome/accessories, disinfect if

applicable, then dry thoroughly and reinstall.

(--> See chapters 7.1 - 7.5).

*) Indication in instruments with specimen cooling. Instruments without specimen cooling: only CT (centered) is displayed.

Temperature upon

switching on

≥ 0 °C

8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8

               
               

C M 3 0 5 0 S V s . X X . X X

E N G L I S H

C U T T I N G M O T O R

S L O

W

P O W E R F A I L U R E

1 2 : 0 0

C T . . . . ° C O T . . . . ° C

H

O

M

E

C T . . . . ° C O T . . . . ° C

* )

D R Y M I C R O T O M E

C T . . . . ° C O T . . . . ° C

* )

C T . . . . ° C O T . . . . ° C

* )

Advertising