1 selecting a sectioning mode, Start and stop sectioning, Operation – Leica Biosystems CM3050 S User Manual

Page 35

Advertising
background image

35

Leica CM3050S – Cryostat

Start / stop motorized sectioning
Sectioning mode 1 - Intermittent stroke:
• Simultaneously press and hold RUN/STOP

and RUN/ENABLE or depress and hold foot-

switch.

• To stop the sectioning stroke, release the

keys or the footswitch:

The specimen head stops immediately.

Sectioning mode 2 - Continuous stroke:
• Simultaneously press and release RUN/

STOP and RUN/ENABLE or depress and re-

lease footswitch.

• To stop sectioning:
Press and release RUN/STOP or RUN/EN-

ABLE or depress and release the footswitch:

The specimen head stops upon reaching the

lower limit position.

Sectioning mode 3 - Single stroke:
• Simultaneously press and release RUN/

STOP and RUN/ENABLE or depress and re-

lease footswitch.

• The instrument carries out one complete

section / retraction stroke and the specimen

head stops upon reaching the lower limit po-

sition.

While sectioning is in motion, LED ‘RUN’ is illu-

minated.

While the sectioning motor is deactivated, LED

‘STOP’ is illuminated.

Both LEDs are located in functional area C.

5. Operation

5.4.1 Selecting a sectioning mode /

start and stop sectioning

Three different sectioning modes can be select-

ed for motorized sectioning:

Sectioning mode 1 - Intermittent stroke:

The specimen head travels up and

down as long as the RUN/STOP and

RUN/ENABLE keys are pressed simulta-

neously or as long as the footswitch is

depressed.

Sectioning mode 2 - Continuous stroke:

The specimen head travels up and

down until it is stopped by pressing

RUN/STOP or RUN/ENABLE or by de-

pressing the footswitch.

Sectioning mode 3 - Single stroke:

The specimen head carries out one sec-

tioning stroke and then stops in the low-

er limit position.

Activating a sectioning mode

Press one of the three sectioning mode keys

described above. While a sectioning mode is

active, the LED in the corresponding key will be

illuminated.

Functional areas A and C - Instruments with sectioning motor only