3 selecting sectioning speed, 4 emergency stop, Operation – Leica Biosystems CM3050 S User Manual

Page 37

Advertising
background image

37

Leica CM3050S – Cryostat

5.4.3 Selecting sectioning speed

• Sectioning speed is selected via the sliding

potentiometer. Speed can be changed while

sectioning is in motion.

Caution when working in continuous mode:
If the slider is set to ‘0’, the sectioning motion

stops, however, the actual sectioning mode is

still active - just operating at ‘0’ speed.
If the slider is accidentally moved, sectioning

will be resumed! - Risk of injury!
Therefore:

Do not interrupt sectioning / trim-

ming by setting the sliding potentiom-

eter to zero speed!

See also safety instruction in Chapter

2.3.6 ‘Motorized sectioning’.

Vmax key

As long as the button is pressed, the

sectioning motion is carried out at maxi-

mum speed (210 mm/s).

While Vmax is active, the sectioning window is

inactive.
When the Vmax key is released, the sectioning

window becomes active again.

5. Operation

5.4.4 Emergency stop

• Activate by:

• either pressing the emergency stop

button forcefully (must lock in posi-

tion),

• or depressing the footswitch force-

fully.

Either way, sectioning is interrupted at once.
In both cases, the red

LED in functional

area A lights up, indicating that the emergency

stop function has been activated. Simultane-

ously all 3 sectioning modes are deactivated

(= all 3 LEDs off).

• To deactivate the emergency stop function,

rotate the button in the direction of the ar-

row until it unlocks and slips upward into its

original position.

As soon as the emergency stop button has

been unlocked, the red

LED is extin-

guished.

• If the emergency stop function has been ac-

tivated with the footswitch, unlocking is not

necessary.

As soon as the footswitch is released, the red

LED is extinguished.

• To continue work, select one of section-

ing modes 1 - 3 and restart the sectioning

motion.

See also chapter 2.2.3 ‘Safety - Emergency stop

function’ for further information on the emer-

gency stop.

Functional areas A and C - Instruments with sectioning motor only