42

Instructions for Use V 1.7 RevC - 08/2014

5. Operation

5.5.4 Inserting specimen discs into the

specimen head

• Lock handwheel in upper position.

If a knife has already been inserted

into the knife holder, cover knife edge

with knife guard!

• Loosen clamping screw (3).
• Insert specimen disc in opening (2) (rear

surface of specimen disc must fit tightly

against surface of specimen head).

• Tighten clamping screw (3).

5.5.5 Specimen orientation

• Open clamping lever (4) --> (4a).
• Orient specimen via orienting screws (1)

and (5).

• Retighten clamping lever (4).

Thermoblock
The thermoblock makes it easy to remove fro-

zen specimens from the specimen discs.

Do not store the thermoblock in-

side the cryochamber - store at room

temperature and only place into the

chamber for removing specimens from

discs.

• Place cap (9) on one of the two ends so that

the correct size opening (7) for the disc to

be inserted is exposed.

• Insert pin (6) of specimen disc (5) into the

opening (7).

• After about 20 seconds remove the fro-

zen specimen from the specimen disc (use

tweezers).

• If cap (9) fits too loosely, readjust with

screw (10). Do not tighten screw too much!

• Once the specimen has been detached from

the disc, remove the thermoblock from the

cryochamber.