5.6 Finishing work
5.6.1 Daily routine steps

Remove all specimens from the cryo-

chamber and store safely! See also

corresponding safety instructions in

chapter 2.3.7 ‘Defrosting / handling fro-

zen tissue’!

• Lock handwheel.
• Remove knife from knife holder, disinfect,

dry (see chapter 7 ‘Disinfection, cleaning

and maintenance’) and store in the knife

case.

• Use a cold brush to sweep frozen sectioning

waste into the section waste tray.

• Empty section waste tray (dispose of sec-

tioning waste according to the applicable

regulations on disposal of special category

waste!).

• If necessary, carry out a complete disinfec-

tion procedure (see chapter 7 ‘Disinfection,

cleaning and maintenance’).

• Remove all specimens from the chamber

and store safely.

• Close sliding window.
• Switch off cryochamber illumination.
• Use lock/unlock key to lock display in con-

trol panel 1 (Display background dark =

locked):

5.6.2 Switching off the instrument for a

longer period of time

If you do not plan to use the instrument for an

extended period of time, it can be switched off.
However, since this involves a lot of work

(see below), it should be considered carefully

whether switching off the instrument is really

the better option.

If the instrument is switched off, the microtome

must be removed. Then the microtome, cryo-

chamber and all accessories must be disinfect-

ed and dried!

• For that purpose, follow the instructions in

chapters 7.1 through 7.5.3.

C T . . . . ° C O T . . . . ° C

1 2 : 0 0

Do not switch off mains switch / auto-

matic cutout (see chapter 5.2 ‘Switch-

ing on / fuses’)!