7. Disinfection, cleaning and maintenance

7.1 Safety instructions on disinfection

and cleaning

When disinfecting the instrument,

make sure to observe the following

safety instructions:

• ‘Disinfection and Cleaning’

(Chapter 2.3.10)

• ‘Handling microtome knives/blades’

(Chapter 2.3.4)

• ‘Defrosting / handling frozen tissue’

(Chapter 2.3.7)

• ‘Frozen instrument parts and acces-

sories’ (Chapter 2.3.8)

• ‘Infectious / radioactive material’

(Chapter 2.3.9)

7.2 Spray disinfection with Leica Cryo-

fect

For practical spray disinfection, we recommend
Leica Cryofect which is fully effective at tem-
peratures in the negative range down to -20 °C.

• Set chamber temperature -20 °C or warmer.
• Remove specimens from the chamber and

store safely.

• Carefully remove knife / blade from knife

holder.

• Remove section waste tray from cryocham-

ber and empty.

• Remove remaining section waste from cryo-

chamber.

• Working accessories (brushes, specimen

holders etc.) can be disinfected inside the
cryochamber.

• Once the selected chamber temperature has

been reached, spray Leica Cryofect on the
surfaces and accessories to be disinfected.

• Allow to soak for 15 minutes.
• Use paper towels or another type of dispos-

able cleaning cloths to wipe up disinfectant
and remaining section waste.

• Dispose of paper towels together with sec-

tion waste.

• If previously altered, set chamber tempera-

ture back to its former value.

• Put specimens back into the cryochamber.