Leica CM3050S – Cryostat

7.5 Removing / reinstalling the micro-

tome

7.5.1 Removing the microtome

When removing the microtome make

sure to observe the following safety in-

structions:

• ‘Removing / reinstalling the micro-

tome’ (Chapter 2.3.11)

• ‘Defrosting / handling frozen tissue’

(Chapter 2.3.7)

• ‘Infectious / radioactive material’

(Chapter 2.3.9)

• ‘Disinfection and cleaning’

(Chapter 2.3.10)

• Prior to removing the microtome, carry out

preparation steps as per chapter 7.3 (switch

instrument off, remove accessories etc.)

7.5.2 Disassembling the specimen head

(instruments w/specimen cooling

only)

• Use Allen key size 5 to remove screw (2)

from the center of the specimen head (1).

• Do not twist the refrigerating tube (see

chapter 2.3.11 ‘Removing / reinstalling the

microtome’).

• Hold hook (4) (part of standard delivery) in

one of your hands and with your other hand

rest refrigerating tube (3) on hook.

• Insert hook (4) with refrigerating tube (3) into

the opening above mark ().