7.5.3 Reinstalling the microtome

When reinstalling the microtome

make sure to observe the following

safety instructions:

• ‘Removing / reinstalling

the microtome’

(Chapter 2.3.11)

• Display message ‘Dry Microtome’

(Chapter 2.3.12)

• Lock handwheel in lower position.

• Place microtome in chamber.

• Put one drop of cryostat oil per surface on

the contact surfaces of plastic coupling (11)

and handwheel coupling (12).

• Place plastic coupling (11) on handwheel

coupling (12).

• Push microtome slightly towards the right

until axle (9) fits tightly into coupling (11).

• Unlock handwheel.

• Adjust the parts into each other by rotating

the handwheel slightly backward and forth.

• Lock handwheel again in lower position.

• Tighten screws (8).

• Reconnect plugs of coarse feed motor (6)

and temperature sensor (7).

• Mount specimen head (in reverse order as

described on previous page).

• Insert accessories (storage shelf, section

waste tray etc.).

• Insert sliding window.

• Connect instrument to mains and switch

back on.

7.

Disinfection, cleaning and maintenance

• Remove plug of temperature sensor (7).

• Remove plug of coarse feed motor:

Pull bushing backward (arrow 1) and hold.

Place your thumb between plug (6) and

socket at the microtome (arrow 2).

To remove plug (6) from microtome, pull it

away from the microtome while at the same

time carrying out a slight jogging movement

(arrows 3).

• Loosen screws (8) with Allen key.

• Lift microtome a little and pull slightly to the

left.

• This loosens the plastic coupling (11) which

connects the two axles.

• Remove the microtome from the chamber.