Leica CM3050S – Cryostat

Removal
• The lamp (3) is located behind the glare

shield (4).

• Carefully hold lamp with both of your hands.

Slightly pull the lamp downward at the left

side to remove it from fastening clip (5).

• Pull the lamp towards the left to remove it

from its socket (6).

Installing the replacement lamp
• Hold the lamp in correct position (see draw-

ing below) and push towards the right until it

locks into the socket.

• Push the lamp slightly upwards until it locks

into the clip.

• Put the sliding window back in place.

• Reconnect the instrument to mains and

switch it back on.

7.6

Exchanging the fluorescent light

lamp

Prior to exchanging the lamp, please

read and observe the following safety

instructions:

• ‘Maintenance’

(Chapter 2.3.13)

Removing the sliding window
• Hold sliding window (1) at grip bar (2), lift

slightly and pull out towards yourself.

7. Disinfection, cleaning and maintenance