background image

60

Instructions for Use V 1.7 RevC - 08/2014

9. Appendix

9.1.4 Installation

• Open the shipping box and remove all packing material including the Sty-

rofoam pieces.

• Take the height adjustment device out of the box.
• At the place of installation, place the U-frame on the floor with the open

side facing yourself. Adjust the U-frame parallel to the wall, maintaining a

distance of approx. 10 cm between wall and U-frame.

• Push the (up) button to lower the height adjustment device to the lowest

possible position.

• Plug the height adjustment device into the

mains socket. The unit is now ready for use.

• In order to be able to move the instrument,

use a fork wrench and unscrew the roll-

ers on which the Leica CM3050 S moves by

about 2 cm.

• Roll the cryostat onto the U-frame as far as it

will go. The rear of the cryostat housing has

to fit closely to the rear bar of the U-frame.

• Roll the cryostat onto the U-frame as far as it

will go. The rear of the cryostat housing has

to fit closely to the rear bar of the U-frame.

When placing the instrument onto the

U-frame, make sure the power cord

does not get caught and thus crushed

between U-frame and cryostat housing.