Instructions for use, 6v3, RevB – 01/2013

1

2

3

7.6.3 Adjusting the knife angle

Tools required:

Allen key (size 5).

Proceed as follows:

1. Comply with the safety instructions provided in chapter 7.6.

2. Loosen screws (1).

3. Rotate knife holder (2) selecting the desired knife angle on scale (3)

(recommended angle = 20°).

4. Tighten screws (1).

7.