3 adjusting the knife angle, Adjusting the knife angle – Leica Biosystems CM3600 XP User Manual
Page 102
Advertising
102
Instructions for use, 6v3, RevB – 01/2013
1
2
3
7.6.3 Adjusting the knife angle
Tools required:
•
Allen key (size 5).
Proceed as follows:
1. Comply with the safety instructions provided in chapter 7.6.
2. Loosen screws (1).
3. Rotate knife holder (2) selecting the desired knife angle on scale (3)
(recommended angle = 20°).
4. Tighten screws (1).