2 inserting the specimen stage, non-orientable, Inserting the specimen stage, non-orientable, Operating the leica cm3600 xp – Leica Biosystems CM3600 XP User Manual

Page 94

Advertising
background image

94

Instructions for use, 6v3, RevB – 01/2013

7.5.2 Inserting the specimen stage, non-orientable

Tools required:

Allen key (size 6).

Proceed as follows:

1. Use the knee lever to move the sledge to the right-side reversal point.

For further information, refer to chapters 3.4.2 and 6.4.9.

2. Comply with the safety instructions provided in chapter 7.5.

3. Open the window. For further information, refer to chapter 3.2.3.

4. Remove screw (1).

5 Lift off T-piece (2) and support ring (3).

6. Remove screws (4) and washers (5) located on the right and left side

of the sledge.

7. Insert specimen stage, non-orientable (6) and orient by eye.

8. Put washers (5) in place and tighten screws (4) on the right and left

side of the sledge.

9. Check specimen stage, non-orientable (6) for correct orientation and

proper tightening.