6.8.3 Error codes: meaning and troubleshooting

Error code

Error message

Trouble shooting

Error 01

Error 02

Error 03

Error 04

Error 05

Error 06

Error 07

Error 08

Error 09

Error 10

Error 11

Knife holder – limit position
exceeded.

Limit position of sledge exceeded

Chamber – Alarm temperature
exceeded

Power failure

Data transfer failed (happens rather
frequently subsequent to a power
failure)

Emergency stop activated

Stepper motor – step error

Refrigerating system – pressure
exceeded

Flap limit switch malfunction

Call Technical Service

Re-initialize instrument. If the problem
recurs, call Technical Service.

Re-initialize instrument. If the problem
recurs, call Technical Service.

Check whether there has been a power
failure!

Check the selected alarm temperature
value. It must differ at least 5 K from the
chamber temperature selected.

Check the cable connection between PC
and control unit.

If problem recurs, call Technical Service

Not defined

Deactivate emergency switch

Call Technical Service.

Check the air circulation between the
condenser and the control unit.

Check the distance of the instrument to
the wall (at least 50 cm)!

If problem recurs, call Technical Service

Re-initialize instrument. If the problem
recurs, call Technical Service.

Call Technical Service.