3 initialization – Leica Biosystems CM3600 XP User Manual

Page 38

Advertising
background image

38

Instructions for use, 6v3, RevB – 01/2013

6.3

Initialization

To be able to work with the instrument, after completion
of the start and log-in procedure the microtome must be
initialized.

For initialization, proceed as follows:

1. In the "Main window", click on the INIT button blinking in yellow color

(in "Mode" field).

The "Safety check - initialization" window will open up.

2. Check whether you can confirm each safety query - if not, carry out

the required actions. Subsequently, confirm each safety query by
mouse click on the corresponding check box.

3. Having confirmed all safety queries, click on OK.

The microtome will then be initialized:

The knife holder moves to the uppermost position.

The microtome sledge moves to the end positions on the right and left.

6.

Software

Advertising