Leica CM3600 XP – Cryomacrotome

7.7

Setting the trimming parameters

Proceed as follows:

1. On the computer screen, open the "Main window". Select the desired

section thickness, sledge speed, cutting window and sectioning
program. For further information, refer to chapter 6.4.

For reasons of safety and accuracy, the maximum trimming
thickness should not exceed 100 μm per section.

When trimming biological material (particularly bone),
section thickness should not exceed 50 μm.

2. Configure "Automatic" mode, then press the "Start" button. For further

information, refer to chapter 6.4.

The trimming procedure will be carried out automatically.

7.

Operating the Leica CM3600 XP