2 electrical connections, 3 other connections, 2 electrical connections 4.3 other connections – Leica Biosystems CM3600 XP User Manual

Page 27: Site requirements

Advertising
background image

27

Leica CM3600 XP – Cryomacrotome

4.2

Electrical connections

Country:

Europe

USA

Japan

Installation cord cross section:

2.5 mm

2

2.5 mm

2

2.5 mm

2

Fuse/connection:

16 A/380 V

25 A/208 V

25 A/200 V

Power cable length is 3 m

Leakage current > 3.5 mA

A second protective conductor terminal is located on the instrument.
Both terminals must be used.
There must not be any differences in potential between the two
protective conductor terminals.

4.3

Other connections

Telephone connection
For remote failure diagnosis, a telephone connection must be available to
hook up the modem. The phone connection must be direct, i.e. must not
be controlled by an additional computer (server).

Compressed air supply
If you want to use the optional extraction system:
Compressed air supply for the extraction system (p = 5 - 8 bar) must be
available.