Installation – Leica Biosystems CM3600 XP User Manual

Page 31

Advertising
background image

31

Leica CM3600 XP – Cryomacrotome

5.

Installation

Important!
The connection of an alarm to the plug provided must be
carried out by trained personnel if the voltages are greater
than 33 V AC or 70 V DC.

The external alarm is triggered, when the chamber temperature exceeds
the selected value.

When the temperature dips back down below the defined threshold, the
external alarm is immediately switched off.

Loss of mains power supply (instrument switched off or power failure) will
also trigger the external alarm.

On return of mains supply the alarm is automatically switched off.

The instrument has a socket via which an external alarm can
be output (see Fig. on p. 29 and wiring diagram). The maxi-
mum permitted voltage that can be switched is 250 V.
Because a compatible plug is provided with the instrument,
the user must make a corresponding connection onsite.

Connection for external alarm
Forwarding an external alarm.

Wiring the connection for external alarm in the instrument

Socket
CA 6

21

24

22

Rel

-g1

1

5

3

The wiring diagram shown here illus-
trates how the socket for the external
alarm is wired in the instrument.

The specified digits correspond to those
on the plug provided.