First use of the instrument, 1 connection to mains power, 2 installing the accessories – Leica Biosystems CM1100 User Manual
Page 13: Connection to mains power, Installing the accessories
13
Leica CM1100 – Cryostat
7.2
Installing the accessories
•
Remove the lid.
•
To release, turn the clamping lever (1) of the micro-
tome base plate counterclockwise, fully insert the
knife holder (2) on the base plate from the right and
clamp by relocating the lever (1).
•
Place the section waste tray (3) between the mi-
crotome and knife holder from the side.
•
Place the section waste tray (4) between the mi-
crotome base plate and chamber wall at front.
•
Place the quick-freeze shelf (5) over the handwheel
shaft between the microtome and right-hand cham-
ber wall.
•
Place the brushes in the location holes (6) of the
quick-freeze shelf.
•
Place all tools needed for specimen preparation in
the cryochamber for precooling.
•
Replace the lid.
7.1
Connection to mains power
Please refer to chapter 4 'Technical data'!
During the start-up of the compressor the
nominal voltage must not drop below the val-
ues specified in the 'Technical data'.
Please note that the compressor requires a
start-up current between 10 and 25 A.
Therefore, the electric circuit at the installa-
tion site must be inspected by an electrical
engineer to ensure that it meets the require-
ments for a smooth operation of the instru-
ment.
A constant adequate power supply to the in-
strument must be ensured at all times.
Failure to comply with the above will cause
severe damage to the instrument.
The instrument must be connected to a
grounded mains power outlet socket.
•
The electric circuit at the place of installation has
to be protected separately.
•
Do not connect any other consumers to this elec-
tric circuit.
Caution:
240 V units may be operated only with the step-
up transformer supplied with the instrument!
Failure to comply with this will cause severe
damage to the instrument!
7.
First use of the instrument
Fig. 13.1
2
1
Fig. 13.2
4
3
5
6