Trouble shooting – Leica Biosystems CM1100 User Manual

Page 26

Advertising
background image

26

Instruction manual, V1.2 – 06/2006

9.

Trouble shooting

-

Refreeze specimen onto the disc.

-

Check disc clamping.

-

Check clamping.

-

Refreeze specimen onto the disc.

-

Increase section thickness; reduce
specimen surface area if neces-
sary.

-

Use different part of cutting edge,
or replace blade.

-

Store all tools in the cryochamber.

-

Raise plate when aligning.
Be more careful next time.

-

Select correct temperature.
Wait until the correct temperature
has been reached.

-

Remove ice.

-

Adapt speed.

-

Check clamping.

-

Check clamping.

-

Apply cryocompound on warm
disc; mount specimen and freeze.

-

Use different part of cutting edge,
or replace blade.

-

Dry microtome thoroughly.

-

Cool down anti-roll plate or
reposition plate.

-

Remove static electricity.

-

Remove fat with alcohol/acetone.

-

Remove rust.

-

Specimen insufficiently frozen onto
the specimen disc.

-

Specimen disc not clamped tightly.

-

Blade not clamped tightly enough.

-

Specimen has been sectioned too
thickly and has detached from the
disc.

-

Very hard, inhomogeneous speci-
men.

-

Blunt cutting edge.

-

Brush, forceps and/or cloth too
warm.

-

Plate too high above the cutting
edge. Adjustment was carried out
in direction of the cutting edge.

-

Temperature incorrect for the
tissue to be cut.

-

Ice buildup on blade back.

-

Handwheel speed not uni-
form.

-

Blade not clamped tightly enough.

-

Specimen disc not clamped tightly.

-

Cryocompound applied to cold
specimen disc; specimen detached
from the disc after freezing.

-

Blunt cutting edge.

-

Microtome not properly dried
before reinstallation.

-

Anti-roll plate is too warm or
incorrectly positioned.

-

Static electricity.

-

Fat on the corner or edge of the
anti-roll plate.

-

Rusty blade.

Chatter during sectioning

Condensation on anti-roll plate and
blade during cleaning

Anti-roll plate damaged after
adjustment

Thick-thin sections

Tissue sticks or crumbles on the anti-
roll plate

Problem

Possible causes

Remedies