Leica CM1100 – Cryostat

I

wrong: anti-roll guide positioned too low

II

wrong: anti-roll guide positioned too high

III

right: anti-roll guide correctly positioned

In position III the anti-roll plate is correctly adjusted.

It is recommended to first preadjust it at a great sec-
tion thickness (e.g. 10 μm) and then approach the re-
quired section thickness reducing the value in 5 μm
increments and slightly readjust the anti-roll plate in
height with the knurled knob (9, Fig. 20.2).

The glass anti-roll guide is resistant to ac-
etone.
For cleaning the knife holder, domestic deter-
gents can be used.

8.

Daily operation

•

To raise the anti-roll plate (to move it towards the
blade), turn the knurled knob (9, Fig. 20.2) counter-
clockwise.

•

To lower the anti-roll plate (to retract it from the
blade), turn the knurled knob (9, Fig. 20.2) clockwise.

If the anti-roll plate is positioned incorrectly to the blade
edge, problems I and II will occur. Position III shows
the correct position of the anti-roll plate.

8.2.7 Replacing the glass anti-roll plate

Fig. 21.2

•

Remove the anti-roll guide from the blade by turn-
ing it to the left.

•

Unscrew the knurled knob (4).

•

Remove the white plastic washer (6) and pull the
anti-roll plate with the shaft up out of the swivel arm.

Install the new anti-roll plate:

•

Insert the shaft (2) of the glass anti-roll plate (1) in
the hole of the swivel arm (3) from above ensuring
the pin (6) is located in the notch (7).

•

Place the white plastic washer (5) on the shaft (2)
from below.

•

Screw the grip (4) on to the shaft (2) from below.

After installation, the anti-roll guide needs to be read-
justed as described at 8.2.6.

1

2

4

3

5

6

7

III

I

II

Fig. 21.1