Technical data

Operating temperature range: 18 °C to 35 °C.
All specifications related to temperature are valid only for an ambient temperature up to 22 °C and for an air
humidity lower than 60%!

Type

CM 1100-1

CM 1100-3

CM 1100-11

Nominal voltage

230 V AC ±10%

120 V AC ±10%

100 V AC ±10%

Nominal frequency

50 Hz

60 Hz

50/60 Hz

Power draw

during refrigeration

170 VA

170 VA

170 VA

during defrosting for 10 min.

320 VA

400 VA

650 VA

Maximum start-up current for 5 sec.

10 A

eff.

25 A

eff.

25 A

eff.

Protective class

I

I

I

Mains fuse

2-pole, T 10 A

2-pole, T 10 A

2-pole, T 10 A

Pollution degree

2

2

2

Overvoltage installation category

II

II

II

Heat output

during refrigeration

170 J/s

170 J/s

170 J/s

during defrosting for 10 min.

320 J/s

400 J/s

650 J/s

Transformer fuse protection F3

T 4 A

T 4 A

T 4 A

acc. to DIN-IEC 127-II

UL-listed

UL-listed

Refrigeration

Temperature range

0 °C to -30 °C ±10%
at an ambient temperature of 22 °C
and air humidity

_

<

60% when lid is closed;

adjustable in 1 K increments

Defrosting

automatic hot gas defrosting,
defrost time freely programmable; 1 defrost cycle/ 24 h,
manual defrosting

duration: 20 min.

Power draw

226 W

204 W

235 W /269 W

Refrigerating output

1)

193 W

185 W

190 W

Maximum nominal pressure

15 bar

15 bar

15 bar

Safety factor

3

3

3

Refrigerant

170 g ± 5 g R134a *

170 g ± 5 g R134a *

155 g ± 5 g R134a *

Compressor oil

0.4 l alpha 22, Kyodo *

0.4 l alpha 22, Kyodo *

0.4 l alpha 22, Kyodo

*

1)

Rating acc. to ASHRAE:

Evaporating temperature

-23.3 °C

Liquefier temperature

54.4 °C

Ambient temperature

32.2 °C

Microtome

Type

Rotary microtome

Section thickness setting range

0 - 20 μm, continuously adjustable

Division of the scale

2 μm

Horizontal specimen feed

15 mm ± 1 mm

Vertical stroke

46 mm ± 1 mm

Maximum specimen size

36 mm

Cryocabinet

Dimensions (W x D x H)

570 x 780 x 380 mm

Weight (including microtome)

50 kg

* Refrigerant and compressor oil must be replaced by authorized service personnel!