8.
Daily operation
8.2.5 Lateral displacement
If the sectioning results start lacking quality, the knife
holder can be displaced laterally to use another blade
position.
Fig. 20.1
1
Proceed as follows:
•
Release the clamping lever (1) by turning it coun-
terclockwise and move the knife holder including
the blade laterally to the desired position.
•
Retighten the clamping lever (1) by turning it clock-
wise
8.2.6 Adjusting the anti-roll plate
Parallel adjustment of the anti-roll plate and blade
It may be necessary to align the front edge of the anti-
roll plate parallel to cutting edge of the blade.
Proceed as follows:
•
Loosen screw (10), hold the anti-roll guide at the
grip (8) and align the front edge of glass plate par-
allel with the cutting edge of the blade.
•
Retighten screw (10) after alignment.
10
7
9
8