The thermal block (8) facilitates the removal of the fro-
zen specimen from the specimen disc.

Keep the thermal block outside the cryocham-
ber at room temperature. Only place in the
cryochamber for specimen removal.

12.1 Mobile heat extractor

Specimen freezing on the quick-freeze shelf can be ac-
celerated by the additional use of a heat extractor.

12.

Optional accessories

12.2 Thermal block

Place the cap (9) on the required side, so that the
appropriate location hole for the specimen disc is
visible.

Insert the shaft (6) of the specimen disc (5) in the
appropriate location hole (7) at the top or bottom of
the thermal block.

After about 20 seconds, the frozen specimen can
be removed from the specimen disc with forceps.

If the cap is too loose, readjust it with the small
screw (10).
Do not overtighten the screw.

Once the specimen is removed, take the thermal
block out of the cold cryochamber.

Store the heat extractor in the cryochamber.

Place it on the specimen surface to accelerate
freezing.

Remove it once the specimen is entirely frozen.