Leica SP2600 – Ultramilling attachment

Adjusting the miller
The miller (3) is adjusted in the holder (2) along

three different axis (x, y and z - see Fig. 5.3, left).
Adjustment along the x-axis is done outside the

calibration device (18, Fig. 5.1).
For adjustment along the z- and y-axis, the miller

holder (2) with miller (3) has to be fastened in the

calibration device.

Adjustment along the x-axis
• Adjust miller (3) as shown in Fig. (5.4a / 5.4b).
• Along the entire length of both sides of the

miller, there must be a uniform distance be-

tween miller (3) and miller holder (2) / spring

(1) (see Fig. 5.4b).

• The miller is adjusted along the x-axis via the

two regulating screws (6) (Allen key, size 2.5).

• Tighten (clockwise) and/or loosen (counter-

clockwise) screws (6, Fig. 5.3) alternately un-

til the miller is adjusted to a parallel position

as shown in Fig. 5.4b.

Do not tighten screws (6) too much. If

screws (6) are too tight, screw (7, Fig.

5.2) (for adjustment along the y-axis)

can no longer be moved.

Fig. 5.3

5. Operation

X

2

3

6

6

Z

Y

Fig. 5.4a: WRONG

Fig. 5.4b: CORRECT

1 3

2