Operation – Leica Biosystems SP2600 User Manual

Page 26

Advertising
background image

26

Supplementary instruction manual V 1.2 – 07/2002

Adjusting the finishing miller
Just like the pre-miller, the finishing miller is ad-

justed in the miller holder along three different

axis (see page 21).

Adjustment along the x-axis
As described for the pre-miller - see page 21.

Adjustment along the z-axis
• Carry out steps 1 - 4 as described on page 22

(Fig. 5.5) for the pre-miller.

Next:
5. With lever (12), bring contact plate (17) close

to the miller (turn lever backwards) so that only

a small gap remains (see Fig. 5.11).

Caution:

The gauge contact plate must never hit

the milling edge.

This could damage the surface of the

contact plate, leading to incorrect

measuring results.

Be very careful when bringing the con-

tact plate close to the milling edge.

Always use the magnifier.

Fig. 5.11

17

12

5. Operation

Advertising