Operation – Leica Biosystems SP2600 User Manual

Page 24

Advertising
background image

24

Supplementary instruction manual V 1.2 – 07/2002

Adjustment along the y-axis

1. Select reference value ‘1’.

The reference value will be needed

later on to adjust the finishing miller to

the same milling level as the pre-

miller.

Any reference value can be selected.

Experience, however, shows that ‘1’ is

the best value for new millers (i.e. mill-

ers not yet resharpened).

2. To select the reference value, move lever (12),

setting the dial of the inner (smaller) scale (13)

to a position just below ‘1’ and the dial of the

outer scale (14) to approximately ‘70’ (approx.

10 o’clock) (Fig. 5.9a).

3. Rotate regulating screw (7) clockwise, bring-

ing the milling edge (3) close to the contact

plate (17).

4. Slowly continue to turn screw (7) clockwise

until the dial of scale (14) starts to move (i.e.

until the milling edge touches the contact

plate).

5. Keep rotating screw (7) very carefully and

slowly until the dial of scale (14) is in the up-

per ‘0’ position and the dial of scale (13) is lo-

cated exactly at ‘1’ (Fig. 5.9b).

Caution:

Be very careful when bringing the con-

tact plate close to the milling edge.

Always use the magnifier.

Make sure the milling edge does not

hit the contact plate!

Fig. 5.9 a

Fig. 5.9 b

12

14

13

7

17

3

7

14

13

5. Operation

Advertising