8 finishing milling, Finishing milling, Operation – Leica Biosystems SP2600 User Manual

Page 40

Advertising
background image

40

Supplementary instruction manual V 1.2 – 07/2002

5.8 Finishing milling

• Attach the miller holder containing the finishing miller to the milling

spindle (Allen screws / Allen key size 5) - see Fig. 5.21, page 34).

Always slide the plexiglass cover to the upper limit position prior

to attaching the miller holders to the milling spindle!

• Bring the cutting edge of the miller close to the specimen surface and

thoroughly check the distance between miller cutting edge and speci-

men surface as described for the pre-miller on page 35.

• If necessary, adjust the milling window (button (61) and others - see

chapter 5.4.14, page 37 of Leica SM2500 instruction manual).

Setting the milling parameters for the finishing milling procedure
• Once again activate automatic mode of operation ‘AUTO’ and set the

milling parameters using the same control unit buttons as described for

the pre-miller on page 36.

Individual parameter settings depend on specimen material and

size. Please refer to the recommendations below and on the fol-

lowing page.

Recommended milling thickness settings for finishing milling
• 5 µm - keep milling until surface is completely level.
• 3 µm - three entire sledge strokes (milling and return stroke).
• 2 µm - three entire sledge strokes (milling and return stroke).
• 1 µm - three entire sledge strokes (milling and return stroke).

Do not exceed a milling thickness setting of 5 µm with the follow-

ing materials: brittle, very ductile materials such as titanium or fi-

berglass and compound samples containing such materials (such

as printed circuit boards).

5.

Operation

Advertising