Supplementary instruction manual V 1.2 – 07/2002

5.

Operation

5.5.4 Connecting the vacuum pump /

placing specimens onto the vacuum

stages

IMPORTANT:

Prior to first use, every vacuum table

surface must be milled once, just like

a specimen surface. This ensures that

the surface of the specimens milled on

the table later on will be perfectly

even.

To mill a table surface, carry out one

entire pre- and finishing milling cycle

as per chapters 5.6 - 5.8, pages 34 - 43

of this manual.

Connecting the vacuum pump
• Place vacuum pump (make sure the voltage

rating of the vacuum pump is appropriate for

the mains supply in your laboratory!) on the

floor underneath the microtome bench.

• Connect vacuum pump to mains.
• Slide connecting hose (1) (part of standard

delivery of Leica vacuum pumps) onto the

nozzle on vacuum pump (2).

• Connect the other end of connecting hose (1)

to the nozzle (3) of the vacuum stage.

• Place the specimen onto the vacuum stage.

There are different types of vacuum

stages for different specimen sizes -

see next page.

Fig. 5.18

Fig. 5.17

1 2