Leica SP2600 – Ultramilling attachment

Placing specimens onto vacuum stage for 30 -

50-mm lids
• Fasten a lid suited to the size of the specimen

to be milled onto the vacuum stage (slotted

screws are part of standard delivery of

vacuum stage lids - use slotted screwdriver).

• Place the specimen (1) onto the sintered sur-

face (2) of the lid.

• Switch on the vacuum pump.

Placing specimens onto vacuum stages for per-

forated specimen holders
• Fasten a perforated specimen holder (3) suited

to the size of the specimen on the vacuum

stage 76 x 25 mm (4) or onto the vacuum stage

with variable surface (not shown here).

• Place specimen on the perforated area of the

vacuum stage

• Switch on the vacuum pump.

The following is recommended for all

vacuum stages, no matter of which

type and size:

If the specimen is smaller than the sin-

tered or perforated stage surface, the

entire stage surface not covered by the

specimen should be covered with

autoadhesive plastic film.

Placing specimens onto vacuum stage 55 x 100

mm
• Place the specimen directly on the sintered

surface area of the vacuum stage.

(Vacuum stage 55 x 100 mm not shown here).

5. Operation

Fig. 5.19

1

2

Fig. 5.20

3

Fig. 5.21

4