Leica SP 2600 – Ultramilling attachment

7. Cleaning, disinfection and maintenance

7.1 Cleaning and disinfection

Only authorized service technicians may access the internal com-

ponents of the instrument for service and repair!

Cleaning detergents appropriate for the ultramilling attachment -

varnished surfaces:

• mild laboratory detergents (slightly moistened cloth)!

non-varnished metal parts:

• mild laboratory detergents,

• xylene substitutes (slightly moistened cloth)

• alcohol (slightly moistened cloth).

No liquids may enter the interior of the instrument during cleaning!

Cleaning
• Clean the basic instrument Leica SM2500 as per the instructions in chap-

ter 7.1, pages 65 - 67 of the Leica SM2500 instruction manual.

• Clean the Leica SP2600 with a soft, slightly moistened cloth using one of

the detergents listed above.

• Clean diamond millers with a soft cloth moistened with alcohol.

For disinfection of the instrument and accessories use Leica Cryo-

fect or other commercial disinfectants. - Use all disinfectants ac-

cording to the manufacturer’s instructions!

No liquids may enter the interior of the instrument during disinfec-

tion!

When disinfecting, wear appropriate goggles and mask as well as

protective gloves!

Dispose of potentially infectious specimen material as per the

laboratory regulations applicable in the country of operation!

Disinfection
• Remove potentially infectious milling waste / remaining samples and dis-

pose of as per the regulations applicable.

• Disinfect instrument and accessories as per the instructions of the disin-

fectant manufacturer.