25

Leica SP2600 – Ultramilling attachment

5.3 Adjusting the finishing millers in the

calibrating device

• Carry out preparatory steps as per chapter 5.2,

page 20 (‘Adjusting the pre-millers in the cali-

brating device‘).

Twisted or deformed springs for the

miller holders must be exchanged!

• Insert spring and finishing miller as described

in chapter 5.2, page 20 (‘Inserting the miller into

the miller holder’).

When the finishing miller is inserted

into the holder, the letter ‘F’ must be

visible. - That way, the milling edge is

pointing into the right direction.

Fig. 5.10

5. Operation