22

Supplementary instruction manual V 1.2 – 07/2002

Adjustment along the z-axis
1. Pull lever (12) at the left of calibration device

(18) forward as far as it will go, thus moving

contact plate (17) backwards to its rear limit

position. This prevents the contact plate from

hitting the milling edge during the adjustment

procedure.

2. Once miller (3) has been correctly adjusted

along the x-axis, place miller holder (2) with

miller (3) next to block (8) into the calibration

device.

3. Fit the clamping device (9) (with the semicir-

cular recess pointing in your direction) against

the miller holder (2) and fasten with two

knurled screws (10) (turn screws clockwise).

4. Rotate magnifier (16) forward to a position

above the miller.

5. With lever (12), bring contact plate (17) close

to the miller (move lever backwards) until just

a very small gap remains (see Fig. 5.6).

Caution:

The gauge contact plate must never hit

the milling edge.

This could damage the surface of the

contact plate, leading to incorrect

measuring results.

Be very careful when bringing the con-

tact plate close to the milling edge.

Always use the magnifier.

Fig. 5.5

18

12

2 3

8

9

10

16

17

Fig. 5.6

(12)

17