Operation – Leica Biosystems SP2600 User Manual

Page 27

Advertising
background image

27

Leica SP2600 – Ultramilling attachment

Fig. 5.12

4

11

3

7

6 Insert size-5 Allen key into regulating screw

(4) insert size-2.5 Allen key into regulating

screw (11) (Fig. 5.12) .

The arrows in Fig. 5.12 indicate clock-

wise sense of rotation.

7 Tighten/loosen screws (11) and (4) alternately

to adjust miller (3) along the z-axis. The miller

is adjusted correctly when angle ’a‘ equals

angle ’b’, i.e. when the distance between each

end of the milling edge and the contact plate

is the same: in Fig. 5.13a the finishing miller

correctly adjusted.
The miller MUST NOT be positioned as shown

in Fig. 5.13b and 5.13c!

Tightening either regulating screw (4,

11) (= clockwise rotation in the sense

of the arrow) moves the corresponding

side of the miller towards the contact

plate.

8. If one of the two screws has been tightened

as far as it will go, slightly release the oppo-

site screw . This slackens the screw that had

been completely tightened.

Do not tighten screws (4, 11) too much.

If screws (4, 11) are too tight, screw (7)

(for adjustment along the y-axis) can

no longer be moved.

Caution:

Be very careful when bringing the

contact plate close to the milling edge.

Always use the magnifier.

Make sure the milling edge does not

hit the contact plate!

Fig. 5.13 a - CORRECT: angle a = angle b

Fig. 5.13 b - WRONG

Fig. 5.13 c - WRONG

a

b

5. Operation