Instructions for Use V 1.8 RevC – 09/2013

• To lock the handwheel, press the lever (5)

outwards and continue to turn the handwheel

slowly until it locks exactly in the 12 o'clock

position.

2. Safety

2.3 Integrated safety devices

Locking the handwheel
There are two ways of locking the handwheel

(12):
Using the lever (3) on the right side of the micro-

tome base plate, the handwheel can be braked

in almost any position.
• To brake, rotate the lever in a counterclock-

wise direction to position .

3

5

12

Caution!

The braking lever (3) must be exactly

in position , so that the handwheel

brake is applied correctly.

If the lever is moved beyond this point,

it is possible that the handwheel is no

longer braked.

• To unlock the handwheel brake turn the lever

(3) back to its original position. Position

.

When using both brake systems at the

same time, always move the lever (3)

to position first.

Otherwise, it may be not be possible to

release the lever (5).