7. Troubleshooting

Problem

Possible cause

Corrective action

• When using cassettes, particu-

larly thin-walled cassettes, from

a company other than Leica or

Surgipath, the cassette may be-

come deformed or other clamp-

ing problems may arise. If while

trying to clamp the cassette you

realize that it is not securely

clamped into place, an alterna-

tive tensioning clamp must be

used.

• The blade is dull.

• The pressure plate is damaged

or incorrectly adjusted.

• Clearance angle of the knife/

blade too small.

• Laterally displace the knife hold-

er or insert a new blade.

• Insert a new pressure plate or

use a new knife holder.

• Readjust the pressure plate.

• Systematically try several clear-

ance angle adjustments, until the

optimum angle is found.

2. Section compression

The sections are very com-

pressed, show folds or are

squeezed together.

• The blade is dull.

• The specimen is too warm.

• The sectioning speed is too fast.

• Use another area of the blade or

a new blade.

• Cool the specimen before sec-

tioning.

• Reduce the sectioning speed.

3. "Stripes" in sections

For knife holder E

• There is an accumulation of par-

affin on the rear pressure plate

of the knife holder.

• Remove paraffin from this area

on a regular basis.