4 assembling the handwheel, Assembling the handwheel, Instrument setup – Leica Biosystems RM2235 User Manual
Page 18
Advertising
16
Instructions for Use V 1.8 RevC – 09/2013
4.
Instrument Setup
The feather key (4) is loosely
placed in the handwheel shaft (1)
and fixed in place with a cable tie
during transport.
• Remove the cable tie (3).
Caution!
Make sure not to lose the
feather key!
• Place the handwheel (2) on the
handwheel shaft (1) as shown.
• Tighten the screw (2a) located
in the center hole of the hand-
wheel with a Allen key No. 4 (5).
• Remove the cover foil from the
self-adhesive cover disk and
fix the cover disk on the hand-
wheel.
4.4 Assembling the handwheel
Fig. 9
1
3
2
2a
5
4
The handwheel has to be assembled before attempting
to use the instrument.
The necessary parts and tools can be found in the toolkit
supplied in the delivery.
Advertising