2. Safety

Safety instructions — working with the instrument

• Always lock the handwheel before cleaning!
• Do not use any solvents containing acetone or xylene for cleaning!
• Ensure that liquids do not enter the interior of the instrument during cleaning!
• When using cleaners, please comply with the safety instructions of the manufacturer

and the laboratory safety regulations!

Hazards — servicing and cleaning

• Prior to manipulating the knife and specimen, or changing the specimen or knife, and

during breaks, always lock the handwheel and cover the cutting edge with the knife

guard.

• ALWAYS turn the handwheel clockwise; otherwise, the brake will not work properly.
• Always wear safety goggles when sectioning brittle specimens. Specimens may splin-

ter!

• Specimen blocks must NOT be oriented during the retraction phase. If a block is oriented

during retraction, the block will advance by the retraction value PLUS the selected sec-

tion thickness before the next section. This may cause damage to both specimen and

knife!

• Prior to sectioning, check that the specimen is securely clamped in the specimen clamp

– failure to observe this poses the risk of damaging the specimen.