Cleaning and maintenance – Leica Biosystems RM2235 User Manual

Page 70

Advertising
background image

68

Instructions for Use V 1.8 RevC – 09/2013

• Lay the removed parts on an absorbent cloth into

the drying chamber (up to a max. of 65°C) and let

the paraffin contamination run off.

8.

Cleaning and Maintenance

Knife holder E

10

11

9

Fig. 93

86

83

83

86

87

11

10

If several knife holders are cleaned at the

same time, the parts must NOT be mixed

up! Failure to adhere to this may result in

sectioning problems!

Take the knife holder apart for cleaning. To do so,

proceed as follows:

• Fold knife guard (9) downward.

• Rotate the clamping lever (11) of the lateral

displacement forwards and pull it out sideways.

• Push the knife holder base plate (86) with the

pressure plate (83) until it can be removed from

the segment arch (87).

• Rotate the blade clamping lever (10) downward

and pull it out sideways.

• Remove pressure plate (83).

• Clean all parts of the knife holder.

There is a burn hazard when remov-

ing the parts from the drying chamber

(65 °C). Wearing safety gloves is rec-

ommended!

For cleaning and removal of paraffin, do

not use xylene or cleaning fluids contain-

ing alcohol (e.g. glass cleaner).

• Dry knife holder and reassemble.

• After cleaning the moving parts, apply a thin coat

of drive part oil to them (

refer also to Chapter 8.3

).

• When installing the pressure plate (83) make sure

that the upper edge is parallel and level with the

rear edge of the knife holder base plate (86).

Fig. 92

Advertising