5 super mega-cassette clamp, Optional accessories – Leica Biosystems RM2235 User Manual

Page 36

Advertising
background image

34

Instructions for Use V 1.8 RevC – 09/2013

6.2.5 Super mega-cassette clamp

Fig. 30

1

2

3

4

5

Assembly of the super mega-cassette clamp

The super mega cassette clamp should preferably be used with the rigid fixture for specimen

clamps.

To do so, proceed as follows:

Remove the rubber ring (5) only after having attached the cassette clamp on the specimen

head!

• Screw the rigid fixture for specimen clamps (4) onto the specimen head (3):
Remove the screw (1), place the specimen holder fixture (4) onto the specimen head (3) from

the front and tighten the screws (2) with an Allen key No. 3.

Next, insert the screw (1) from the side and briefly tighten it with an Allen key No. 4.
• Insert the super mega cassette clamp from the side on the left into the dovetail guide of the

rigid fixture for specimen clamps and tighten screw (1).

6.

Optional Accessories

If the directional fixture for specimen clamps is used with the rigid knife holder base, the

orientation must be in position "0" and the cover for backlighting illumination must be

detached. (Danger of collision if not observed!)
NEVER use the super mega-cassette clamp with backlighting illumination!

Advertising