3 fine adjustment of the force balance, Fine adjustment of the force balance, Operation – Leica Biosystems RM2235 User Manual

Page 25

Advertising
background image

23

Leica RM2235

33

34

Important!

Never turn the

screw more
than

1

/

2

turn at

a time.

5.3 Fine adjustment of the force balance

If another accessory of a different weight is

mounted on the specimen head (33), you must

check whether it is necessary to readjust the

force balance.
Checking the correct setting:

• Attach the new accessory and clamp the

specimen.

• Set the specimen head to half the height of the

vertical travel range by turning the handwheel

(

Fig. 19

).

If the specimen head remains in this exact posi-

tion, the setting is correct.

If the specimen head moves, i.e. it is raised or

lowered, fine adjustment is necessary.

5. Operation

Failure to adjust the force balance may

result in injury while working.

The force balance is adjusted using the screw

(34), which can be accessed by removing the

section waste tray on the bottom of the base

plate of the microtome. Use the Allen key pro-

vided, No. 5 (with handle) for the adjustment.
• If the specimen head moves downwards, turn

the screw approx.

1

/

2

turn clockwise.

• If the specimen head moves upwards, turn the

screw (34) approx.

1

/

2

turn counterclockwise.

• Continue this procedure until the specimen

head no longer moves once released.

Fig. 19