3 lubricating the instrument, Lubricating the instrument, Cleaning and maintenance – Leica Biosystems RM2235 User Manual
Page 72
70
Instructions for Use V 1.8 RevC – 09/2013
8.
Cleaning and Maintenance
97
88
73
11
10
Knife holder E:
• Clamping lever (11) for the lateral displace-
ment.
• T-piece (88) and guide (97) of the knife holder
for the lateral displacement.
• Clamping lever (10) for the blade.
Fig. 95
Knife holders N and NZ:
• Sliding surfaces of the finger guard (8).
• Knurled nuts (73) for measuring height adjust-
ment.
95
96
55
8.3 Lubricating the instrument
Once a month, lubricate the following parts with
the included drive parts oil No. 405 (1 - 2 drops
are enough).
Instrument and specimen holder:
• The clamping key (95) of the quick clamping
system.
• The T-piece (55) on the microtome base plate.
• Guide rails (96) for the knife holder base on the
microtome base plate.
• The T-piece on the knife holder base (56).
56
8
Knife holder N
Knife holder NZ
91