Optional accessories – Leica Biosystems RM2235 User Manual

Page 47

Advertising
background image

45

Leica RM2235

2. Attach vertical column

• Insert the cap screw (b) into the hole of the

baseplate from below. Place the lock washer

(c) on the cap screw from above.

• Thread the silver vertical column (2) onto the

baseplate from above and tighten with the

Allen key No. 8.

a

3. Attach horizontal arm

• Slide the support ring (e) onto the vertical

column and position it so that the lock nut (f)

is facing to the back. Tighten the lock nut.

• Slide the cross-member (d) onto the column.

Ensure that the lock screw (g) is facing to the

right of the baseplate. The horizontal arm must

be centered over the microscope.

• Slide the horizontal arm (3), flat side facing the

lock screw (g), into the cross-member (d) and

tighten.

6. Optional Accessories

Fig. 49

Fig. 48

Fig. 47

b

c

2

e

d

2

g

f

h

For more information on connecting

and using the microscope, magnifying

lens or cold-light source, please see

the appropriate Instructions for Use.

Extreme danger of tipping! After in-

stalling the vertical column, imme-

diately place the microtome onto the

baseplate so that the front feet of the

microtome are positioned in the shal-

low recesses (A).

The lock washer must be positioned

between the baseplate and vertical

column to prevent unintentional turn-

ing of the column.