How to tune a clc control mode, How to tune a clc control mode 107 – MTS Series 793 User Manual

Page 107

Advertising
background image

How to Perform Advanced Tuning Techniques

MTS Series 793 Tuning and Calibration

Tuning Procedures

107

How to Tune a CLC Control Mode

Tune the control modes that will supply the active and limiting feedback signals

.

1. Select an access level of Tuning:

Standalone FlexTest SE Controllers: Select Tuning > P Gain.

Automated Controllers: In the Station Manager window’s toolbar, select
an access level of Tuning.

10.0

Upper

Limit

Lower

Limit

Force

Feedback

Proportional Gain

10.0

Manual

Command

Displacement

Feedback

Servovalve

Command

Upper Limit

Error

Lower Limit

Error

+

+

+

Displacement

Error

Active P Gain

Limiting P Gain

Bandwidth

2.5

Bandwidth

The CLC control mode uses one of three error signals. The Limiting P Gain adjustment
acts as a conversion factor to scale the limit feedback to similar units as the active P
feedback. The Limiting P Gain is turned on within the upper and lower limit bands
defined by the Bandwidth control.

Otherwise use displacement error

If force feedback is moving toward
the lower limit And displacement
error is negative Then use lower
limit error.

Defines bands at the upper and
lower limits within which Limit P
Gain
is applied.

If force feedback is moving toward
the upper limit And displacement
error is positive Then use upper
limit error.

Channel Limited Channel Diagram