MTS Series 793 User Manual

Page 169

Advertising
background image

Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

169

Using a load standard

Adjust Manual Cmd slider for 0 kN. Then zero the load standard
readout.

Using dead weights

Remove all dead weights. On the Inputs panel, click on the Offset/
Zero
tab, and then adjust Manual Offset for a signal value of 0 kN on
the DVM or Station Signals panel.

D. Adjust the Manual Cmd slider to achieve a load standard reading of

20% compression and record the meter reading. Repeat this step for
40%, 60%, 80%, and 100% compression.

2. Record the data points for the tension.

Standalone FlexTest SE Controllers:

A. Adjust the Setpoint to achieve a load standard readout of zero.

B. Adjust the Setpoint between zero and full tension three times. This

exercises the force sensor to remove hysteresis.

C. Establish the zero reference.

Using a load standard

Adjust the Setpoint for 0 kN. Then zero the load standard readout.

Using dead weights

Remove all dead weights.

Adjust Manual Offset for a signal value of 0 kN on the DVM or other
monitoring device.

Select Setup > Force > Offset/Zero > Manual Offset

D. Adjust the Setpoint to achieve a force standard reading of 20% tension

and record the meter reading. Repeat this step for 40%, 60%, 80%, and
100% compression.

Automated Controllers:

A. Adjust the Manual Cmd slider to achieve a load standard readout of

zero.

B. Adjust the Manual Cmd slider between zero and full tension three

times. This exercises the force sensor to remove hysteresis.

C. Establish the zero reference.

Using a load standard

Adjust the Manual Cmd slider for 0 kN. Then zero the load standard
readout.

Using dead weights

Remove all dead weights. Then adjust the Offset control for a Signal
Value
of 0 kN on the Input Signals window.